74 Granit XP User Guide
by this setting. After the maximum number of attempts is reached, the scanner will 
not attempt to reconnect to a base or an Access Point. Pressing the trigger, scan-
ning an Access Point linking barcode, or placing the scanner in the cradle resets 
the attempt count and the scanner will again try to link. 
Scan the Maximum Link Attempts barcode, then scan the number of attempts for 
the setting (from 0-100) from theProgramming Chart, beginning on page 291, 
then Save. Default = 0.
Note: When Auto Reconnect Mode is On, setting Maximum Link Attempts to zero will cause 
the scanner to try to link until the Power Time-Out Timer setting expires. When Auto 
Reconnect Mode is Off, setting Maximum Link Attempts to zero will cause the 
scanner to only attempt linking one time after a trigger press.
Relink Time-Out
Relink Time-Out controls the idle time between relink attempts. An attempt to link 
a scanner to a base or an Access Point typically lasts up to 5 seconds. This is the 
time when the scanner is actually attempting a contact. Relink Time-Out controls 
the amount of time, in seconds, that elapses between the end of one connection 
attempt and the start of the next.
Note: The length of time for an attempt depends on the number of scanners connected to 
a base unit or Access Point. An extra 7 seconds may be required when a connection is 
successful.
Scan the Relink Time-Out barcode, then scan the number of seconds for the set-
ting (from 1-100) from the Programming Chart, beginning on page 291, then 
Save. Default = 3 seconds.
Bluetooth/ISM Network Activity Examples
Default values
When the scanner goes out of range, the scanner repeatedly attempts to connect 
to the base unit or Access Point. Each attempt consists of approximately 5 seconds 
of active time followed by 3 seconds of idle time. After one hour, the scanner pow-
ers off and batch mode data is lost.
Example: Maximum Link Attempts set to 15
Other values at default setting