EasyManuals Logo

Honeywell Midas User Manual

Honeywell Midas
160 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #23 background imageLoading...
Page #23 background image
Midas
®
Gas Detector
4-7
4.3 Sample and Exhaust Tubing Calculations
The following tables show the flow rate, tubing
length, transport time, and maximum pressure and
vacuum at the inlet and exhaust points. Tubing
lengths vary among gases. See Appendix B for
recommended lengths.
Table 4-1. Inlet sample specications
Inlet Sample Specifications:
Maximum
Tubing Length, m (ft)
30 (100) 20 (66) 10 (33) 0
Transport Time (sec), ID 1/8”
1
28 19 10 1
Transport Time (sec), ID 3/16”
63 43 23 1
Sample Point Vacuum
(Negative pressure)
-25.4 cm H
2
O (-10 in H
2
O) Maximum
Flow rate, cc/min.
500 (Flow is constant)
2
Tubing OD, mm (in)
6.35 (0.25)
Tubing ID, mm (in)
3.18 (0.125)
1
Honeywell Analytics recommends thick-wall tubing (1/8” ID)
for best speed of response. Due to its lower surface area, thick-
wall tubing may require less conditioning than thin-wall tubing.
2
The ow rate is electronically maintained at approximately
500 cc/min and may vary within acceptable tolerances.
Table 4-2. Outlet sample specications
Outlet Sample Specifications:
Maximum
Tubing Length, m (ft)
30 (100)
Back Pressure at Exhaust Point
(Excluding tubing)
20.3 cm H
2
O (8 in H
2
O) Maximum
Tubing OD, mm (in)
6.35 (0.25)
Tubing ID, mm (in)
4.76 (0.188)
Note
Honeywell Analytics recommends the use of
Teon FEP (Fluorinated Ethylene Polymer)
tubing to assure proper sample transport.
The properties of Teon FEP make it the
best choice for transporting sample toxic
gases to instruments when compared with the
properties of other similar tubing materials.
If the pressure/vacuum on the inlet/exhaust lines
does not meet the recommended values in Tables
4-1 and 4-2, the following chart describes potential
fault conditions that may be brought on by the
external influences thus resulting in an F81 Flow
Fault.
External Flow Fluctuation Fault 81 Explanation
1
Low
(0-150 cc/min)
None No Midas
®
will auto-adjust
2
Low
(0-150 cc/min)
Yes
(up to 100 cc/
min flow swings)
Likely
Midas
®
changes the flow
gradually. If external flow
changes are large and rapid,
the final reading will be
different from what Midas
®
expected.
Midas
®
will issue
Fault 81.
3
Medium
(~200-450 cc/
min)
Yes and No Yes
Minimum flow rate for the
pump is ~300 cc/min. With
this external flow, the auto-
adjust cannot work because
the pump cannot produce
a flow at this low level. The
problem is worse if there are
flow fluctuations.
4
High
(> 600 cc/min)
Yes and No Yes
Midas
®
cannot reduce the
external flow. Midas
®
will
issue Fault 81 due to high
flow

Table of Contents

Other manuals for Honeywell Midas

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell Midas and is the answer not in the manual?

Honeywell Midas Specifications

General IconGeneral
Power Supply24 VDC
Detection TechnologyElectrochemical, Infrared
Sampling MethodDiffusion
DisplayLCD
CommunicationModbus RTU, 4-20mA
Operating Temperature-20°C to +50°C
Humidity Range0% to 95% RH (non-condensing)
Target GasMultiple gases
Response Time< 30 seconds
Operating Humidity0% to 95% RH (non-condensing)
Outputs4-20mA
Protection RatingIP66
CertificationsATEX, UL, CSA
Detection GasesOxygen (O2), Carbon Monoxide (CO), Hydrogen Sulfide (H2S), Methane (CH4), and others

Related product manuals