EasyManuals Logo

Honeywell Midas User Manual

Honeywell Midas
160 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #50 background imageLoading...
Page #50 background image
Midas
®
Gas Detector
6-8
6.3.2 Calibration Menu Overview ‘
CAL
The calibration menu allows calibration of the
detector zero, span, flow and 4-20 mA. The menu
is simply navigated using the ‘s’ up and ‘t’ down
buttons to select the required submenu and then
using ‘3’ accept button to enter that submenu. The
s’ up and ‘t’ down buttons are used to make any
changes to a selected setting and are confirmed
using the ‘3’ accept button. The ‘X’ cancel button
can be used to exit the submenu and allow selection
of a different submenu, or can be pressed again to
exit to the main set-up, calibration and test menu.
Pressing the ‘X’ cancel button again returns the
detector to normal operating mode.
The table below shows an overview of the
calibration submenus and how they are navigated.
For a detailed step-by-step instruction of how to
change the detector calibration settings refer to
Section 7.2.

Table of Contents

Other manuals for Honeywell Midas

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell Midas and is the answer not in the manual?

Honeywell Midas Specifications

General IconGeneral
Power Supply24 VDC
Detection TechnologyElectrochemical, Infrared
Sampling MethodDiffusion
DisplayLCD
CommunicationModbus RTU, 4-20mA
Operating Temperature-20°C to +50°C
Humidity Range0% to 95% RH (non-condensing)
Target GasMultiple gases
Response Time< 30 seconds
Operating Humidity0% to 95% RH (non-condensing)
Outputs4-20mA
Protection RatingIP66
CertificationsATEX, UL, CSA
Detection GasesOxygen (O2), Carbon Monoxide (CO), Hydrogen Sulfide (H2S), Methane (CH4), and others

Related product manuals