16
Honeywell Titan SCBA, 20132
soon as the  ll hose coupling engages into the RIC 
UAC coupling on the SCBA.
5. Filling is complete when the pressure in the SCBA 
cylinder and in the  ll cylinder are equalized. This will 
take place in approximately one minute.
NOTE
In most cases, the  cylinder  will not be  lled to 
full  service  pressure.  The  resulting  service  life 
of the SCBA will be reduced.
6. Disconnect the  ll hose coupling by pulling back on 
the outer locking sleeve. An audible hissing or pop-
ping will be heard when the  ll hose is disconnected. 
When this occurs, coupling separation has been 
achieved.
7. Replace the dust cover on the SCBA RIC UAC 
coupling.
8. Close the cylinder valve on the  ll system and vent 
the air from the  ll hose. Replace the dust cover on 
the  ll hose.
9. Replace or re ll the cylinder in the RIT Kit for future 
use.
H. Emergency Operation
1. PROBLEM: Restricted or interrupted air  ow 
a. Open the bypass valve by turning the red knob 
on the second stage regulator counterclockwise 
until desired constant air  ow is achieved.
WARNING
Activating  the  bypass  valve  will  rapidly 
deplete  your  air  supply.  Immediately  exit 
to  a  safe  area.  Failure  to  comply  with  this 
warning may lead to personal injury, illness, 
or death.
b. IMMEDIATELY exit to a safe area.
c. Have the SCBA inspected and/or repaired by a 
Honeywell-certi ed repair technician before reuse.
2. PROBLEM: First-Breath-On failure
a. Press the manual override button on the front of 
the regulator to start air  ow.
b. IMMEDIATELY exit to a safe area.
c. Have the SCBA inspected and/or repaired by a 
Honeywell-certi ed repair technician before reuse.
3. PROBLEM: Free  ow
a. If the regulator will not shut off (free  ow) during 
extremely heavy breathing, exhale forcefully. The 
regulator should return to normal  ow.
b. If the free  ow continues, open and close the 
bypass once.
c. If the problem persists, IMMEDIATELY exit to a 
safe area.
d. Have the SCBA inspected and/or repaired by a 
Honeywell-certi ed repair technician before reuse.
4. PROBLEM: First stage overpressurization relief 
valve operates
PRV is designed to vent air to atmosphere when 
the  ll pressure exceeds the service pressure of the 
cylinder. The PRV will reset (close) at a pressure 
below the service pressure of the SCBA. If a high 
pressure (4500 psig)  ll (supply) cylinder is used on a 
2216 psig SCBA, the PRV may open and vent air to 
atmosphere.
WARNING
• NEVER use the RIC UAC for routine fi lling 
of SCBA cylinders. The RIC UAC is intended 
to be used for cylinder fi lling during rescue 
operations only.
•  Never  use  the  Honeywell  RIC  UAC  to  fi ll 
an  SCBA  cylinder  while  the  SCBA  is  being 
worn  unless  there  is  a  compelling  reason 
to  assume  the  risk  of  injury  if  there  is  a 
component failure during the fi ll process.
• Never use the Honeywell RIC UAC to fi ll an 
SCBA  air  cylinder  while  the  SCBA is  being 
worn if the SCBA or the cylinder is suspected 
of having been dropped, exposed to direct 
fl ame impingement, or damaged in any way.
•  The  purity  of  the  RIC  UAC  air  source 
must meet the requirements set forth in the 
Compressed  Gas  Association  Commodity 
Specifi cation  for  Air,  G-7.1,  Type  1,  Grade 
D,  with  a  dewpoint  not  greater  then  -65°F 
(-53.9°C), or less than 24 ppm by volume.
• The relief valve on the fi rst stage pressure 
reducer  is  factory  set.  Tampering  with 
the  relief  valve  may  cause  the  SCBA  to 
malfunction  or  may  result  in  a  decrease  in 
service life.
•  If  at  any  time  during  fi lling,  an  air  leak  is 
detected  or  suspected  in  the  SCBA  or  the 
fi ll  system,  disconnect  the  fi ll  hose  from 
the  SCBA.  Close  the  cylinder  valve  on 
the  fi ll  cylinder  and  vent  the  air  from  the 
fi ll  hose.  Remove  the  SCBA  from  service 
for  inspection  and  repair  by  a  Honeywell-
certifi ed technician before use.
• The dust covers on the RIC UAC coupling 
and the RIC UAC fi ll hose must be installed 
at  all  times  unless  the  SCBA  cylinder  is 
being fi lled via the RIC UAC.
•  Failure  to  comply  with  this  warning  may 
lead to personal injury, illness, or death. 
NOTE
Air venting from the PRV is extremely loud. 
1. Open the cylinder valve on the RIC UAC  ll system 
(supply cylinder). The  ll hose is now pressurized.
2. Remove the dust cover from the RIC UAC  ll hose.
3. Remove the dust cover from the RIC UAC coupling 
on the SCBA.
4. Connect the  ll hose to the RIC UAC coupling on 
the SCBA. Push the female coupling on the  ll hose 
until a click is heard. The cylinder will start  lling as