OTlMe
OTIMe
command/query
The OTIMe command positions the 0 Marker on the timing waveforms in
the mixed mode display. If the data is not valid, the command performs
no action.
The OTIMe query returns the 0 Marker position in time. If data is not
valid, the query returns 9.9E37.
Command Syntax:
:WLISt:OTIMe
<time-value>
where:
c
time value
>
:: = real number
Example:
OUTPUT XXX,“:WLIST:OTIME
40.Oe-6”
Query Syntax:
:WLISt:OTIMe?
Returned Format:
[:WLISt:OTIMe]
c
time-value
>
<
NL>
Example:
10 DIM
To$[lOO]
20 OUTPUT
XXX;“:WLIST:OTIME?”
30 ENTER
XXX;To$
40 PRINT
To$
50 END