XTAG
XTAG
command/query
The XT’AG command specifies the tag value on which the X Marker
should be placed. The tag value is time when time tagging is on or states
when state tagging is on. If the data is not
valid
tagged data, no action is
performed.
The XTAG query returns the X Marker position in time when time
tagging is on or in states when state tagging is on, regardless of whether
the marker was positioned in time or through a pattern search. If data is
not valid tagged data, the query returns 9.9E37 for time tagging, 32767 for
state tagging.
Command Syntax:
:MACHine{
1
I2):SLISt:XlAG
{
<time-value
>
1 <state-value
>}
where:
-z
time-value
5
:: = real number
<state value
>
:: = integer
Example:
:OUTPUT XXX;" :MACHINEl:SLIST:XTAG
40.OE-6"
Query Syntax:
:MACHine{
1/2}:SLISt:XTAG?
f?eturrled Format:
[:MACHine{
1
I2}:SLISt:XTAG]
{
c
time-value
>
1 estate-value
>}
<
NL>
Example:
10 DIM
xt$[loo]
20 OUTPUT
XXX;“:MACHINEl:SLIST:XTAG?”
30 ENTER XXX;Xt$
40 PRINT Xt$
50 END