EasyManua.ls Logo

HP 4262A - Page 100

HP 4262A
264 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model
4262A
Section
IV
Test
Program
7
PERFORMANCE
TESTS
TEST
PROGRAM
7
[PURPOSE]
This
test
verifies
that
4262A
FUNCTION,
FREQUENCY
and
TRIGGER
can
be
controlled
in
character
(ASCII)
code
and
that
the
measured
data
is
printed
in
accord
with
the
format
shown
in
Paragraph
3-67.
[PROGRAMMING]
Annotation
is
omitted.
ort
"ENTER
REMO1E
PROGRAM
CODE
when
using
STRING-ADV
ROM"
;
spc
3
dim
A$[20],88[25]
rem
7
cli
7
ent
"PROGRAM
CODE
?
(as
F2H3T3)",AS
wrt
717,AS5
;wait
1000
trg
717
red
717,35
prt
3%
spc
3
fas
.
OO~
W
e
oo
*
e
[RESULTS]
The
4262A
REMOTE
lamp
lights.
LISTEN
and
TALK
lamps
alternately
light
once.
Calculator
prints
LCR
and
DQ
values.
Confirms
that
4262A
functions
were
correctly
set
(check
the
printed
data).