Page #100 background imageLoading...
Page #100 background image
Model
4262A
Section
IV
Test
Program
7
PERFORMANCE
TESTS
TEST
PROGRAM
7
[PURPOSE]
This
test
verifies
that
4262A
FUNCTION,
FREQUENCY
and
TRIGGER
can
be
controlled
in
character
(ASCII)
code
and
that
the
measured
data
is
printed
in
accord
with
the
format
shown
in
Paragraph
3-67.
[PROGRAMMING]
Annotation
is
omitted.
ort
"ENTER
REMO1E
PROGRAM
CODE
when
using
STRING-ADV
ROM"
;
spc
3
dim
A$[20],88[25]
rem
7
cli
7
ent
"PROGRAM
CODE
?
(as
F2H3T3)",AS
wrt
717,AS5
;wait
1000
trg
717
red
717,35
prt
3%
spc
3
fas
.
OO~
W
e
oo
*
e
[RESULTS]
The
4262A
REMOTE
lamp
lights.
LISTEN
and
TALK
lamps
alternately
light
once.
Calculator
prints
LCR
and
DQ
values.
Confirms
that
4262A
functions
were
correctly
set
(check
the
printed
data).
4-33

Table of Contents

Other manuals for HP 4262A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 4262A and is the answer not in the manual?

HP 4262A Specifications

General IconGeneral
BrandHP
Model4262A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals