Section
I
Model
4262A
Table
1-1
Table
1-1.
Specifications
(Sheet
3
of
4).
L-D,
L-Q
MEASUREMENT
120Hz
|
1000xH
|
10.00mH|100.0mH
[1000mH
|10.00H
|
100.0H
|
1000H
L
1kHz|
100.04H
|
1000¢H
|10.00mH
[100.0mH
|
1000mH
|
10.00H
|
100.0H
Ranges
10kHz
|
10.00xH
[
100.04H
[1000xH
[10.00mH
|
100.0mH|
1000mH
|
10.00H
D
.001~19.9
(2
Ranges)
Q"
0.05~
1000
(4
Ranges)
Lo+
L
1v
]
TestSignal
"
o
|
40mA
_
10mA
ImA
_
100uA
_
104A
|
auto
||
Sameas
Mode
||
Sameas
Tt
Mode
|
(At
120Hz,
1kHz)
[
0.83%+2counts
|[
1%+
2counts
|
'@D"
(At
10kHz)
[
0.3%
+2counts
|[1%+2
]
5%+2
|
L
Accuracy*s
[
0.2%
+
2
counts
|
(At
120Hz,
1kHz)
T
W
[0.3%
+
2]
0.2%
+
2
counts
|
(At
10xHz)
auto
||
Sameas
-w
Mode
||
Sameas
{JY
Mode
|
(At
120Hz,
1kHz)
[0.8%
+
(3
+
Lx/500)[1%
+
(3
+
Lx/500)
-(33}
(At
10kHz)
[0.5%
+
(3
+
Lx/500)|
[
15+
6%
|[
s+
65
|
R((xl:{?gc
L
0.2%
+
(3
+
200/Lx)
counts
—l
(At
120Hz,
1kHz)
y
-
[
0.5%
+
(8
+
200/Lx)
counts
|
(At
1kHz)
AUTO
||
Sameas
-www
Mode
||
Sameas
-@-
Mode
|
*1
Calculated
from
D
value
as
a
reciprocal
number.
*2
Typical
data,
varies
with
value
of
D
and
number
of
counts.
*3
3(%
of
reading
+
counts).
Lxis
inductance
readout
in
counts.
This
accuracy
only
applies
for
D
values
to
1.999.
Accuracy
applies
over
a
temperature
range
of
23°C
%
5°C.
(At
0°C
to
55°C,
error
doubles).
R/ESR
MEASUREMENT
120Hz
Ranges
R/ESRlil)kHZ
1000m$2
[10:008
[100.0Q
|
1000
[10.00k22[100.0k$2|1000k2)10.00MS2
kHz
Enian
[
1v
]
i
—1wW—
.
Test
Signal
40mA
10mA
1mA
100uA
10uA
Level
*1
YA
AUTO
||
Sameas
4bw--mwwMode
|[
Sameas
1wt
Mode
|
o
[
0.3%
+
2
counts
*?
]
-
0.2%
+
2
ts
|
Accuracy
*2
W
2%
coun
AUTO
||___Sameas
4pw-wrw-Mode
|[
Sameas
Gt
Mode
|
*1
Typical
data,
varies
with
number
of
counts.
*2
£(%
of
reading
+
counts).
*3
(0.5%
+
2
eounts)
on
10.00MQ
range
at
10kHz.
*%
Measurement
range
for
ESR
(equivalent
series
resistance)
is
from
1m$
to
19.99kQ
(typical),
which
varies
with
series
capacitance
and
inductance
value
. .
. .
refer
to
“REFERENCE
DATA”.
Accuracy
applies
over
a
temperature
range
of
23°C
£
5°C.
(At
0°C
to
55°C,
error
doubles.)
1-4

Table of Contents

Other manuals for HP 4262A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 4262A and is the answer not in the manual?

HP 4262A Specifications

General IconGeneral
BrandHP
Model4262A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals