Section
IV
Model
4262A
Paragraph
4-12
PERFORMANCE
TESTS
4-12.
DISSIPATION
FACTOR
CONFIRMATION
CHECK
DESCRIPTION:
This
test
verifies
that
a
tested
4262A
satisfies
dissipation
factor
measurement
ac-
curacies.
Only
one
Dissipation
Factor
(D
=
1.8)
is
checked
for
120Hz,
1kHz
and
10kHz
in
this
check
because
only
one
detecting
phase
needs
to
be
checked.
All
other
factors
influencing
D
accuracy
were
checked
in
paragraph
4-10.
Note
Dissipation
factor
accuracy
for
only
one
D
standard
which
has
a
D
value
of
approximate-
ly
1.8
need
be
proved
to
guarantee
D
accura-
cy.
This
test
also
verifies
that
4262A
correctly
calculates
Q
factor
as
a
reciprocal
number
of
Dissipation
Factor.
Only
one
Q
factor
corres-
ponding
to
a
D
value
of
approximately
1.8
is
checked
in
this
fest.
D
accuracy
in
measuring
inductance
does
not
need
to
be
checked
because
detecting
phase
accuracy
is
equated
with
that
for
capacitance
measurement.
C-D
ACCURACY
SPECIFICATIONS
120Hz
|
1000pF|10.00nF{100.0nF|
1000nF|10.004F|100.0xF|
1000uF|10.00mI
Range
1kHz
1100.0pF|
1000pF{10.00nF(100.0nF|
1000nF|10.004F|100.0uF|
1000uF
10kHz
10.00pF
]100.0pF|
1000pF|10.00nF|100.0nF|
1000nF|10.00uF
100.0uFj
[
0.2%
+
(2
+
200/Cx)
counts
]
At
120Hz,
1kHz
(Test
signal
level:
1V)
-Dil}
[
0.5%
+
(2
+
200/Cx)
countd
At
10kHz
[
0.3%
+
(2
+
1000/Cx)
counts
|
At
120Hz,
1kHz
D
(1
(Test
signal
level:
50mV)
A
w1
[[_1.0%
+
(2
+
1000/Cx)
counts
|
Nt
10kHy
(At
120Hz,
1kHz)|__0.3%
+
(2
+
Cx/500)
counts
_|fim--&
(At
10kHz)
L
0.5%
+
(2
+
Cx/500)
counts
“EM
®
»m
5%
+(5
‘%]
AUTO
||
Sameas
Rl
ModejL
Same
as
d4MW-
Mode
—]
1w
*1
£(%
of
reading
+
counts). Cx
is
capacitance
readout
in
counts.
Accuracy
applies
over
temperature
range
of
23°C
+5°C.
(At
0°C
to
55°C,
error
doubles)
This
accuracy
only
applies
for
D
values
to
1.999.
DUT BOX
(a)
(b)
Figure
4-4.
Dissipation
Factor
Accuracy
Test
Setups.
4-14

Table of Contents

Other manuals for HP 4262A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 4262A and is the answer not in the manual?

HP 4262A Specifications

General IconGeneral
BrandHP
Model4262A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals