EasyManuals Logo

HP 7683 User Manual

HP 7683
38 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
Installation Guide
Replacing the needle support assembly
18
Replacing the needle support assembly
Use the standard needle support assembly for all injections except cool on-
column injections with a 250-µm or 320-µm column. For these injections, you
must change the needle support assembly to HP accessory 18599T.
Figure 11 Needle support assemblies
To replace the needle support assembly:
1. Remove all vials from turret, and disconnect the injector cable from the GC
and lay the HP G2613A Injector on its back on a flat surface.
2. Open the injector door.
3. Remove the syringe.
530
µ
m
250
µ
m/ 320
µ
m
(standard)
(HP accessory 18599T)
230
µ
m/ 320
µ
m
Needle support foot
Needle support foot

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 7683 and is the answer not in the manual?

HP 7683 Specifications

General IconGeneral
BrandHP
Model7683
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals