Temperature
The
HP
8153A
may
be
operated
in
temperatures
from
O^C
to
55“C.
Humidity
The
HP
8153A
may
be
operated
in
environments
with
humidity
up
to
95%
(O^C
to
+40°C).
The
HP
8153A
should
be
protected
from
temperatures
or
temperature
changes
that
cause
condensation
within
the
instrument.
Instrument
Cooling
The
HP
8153A
has
a
cooling
fan
mounted
internally.
Mount
or
position
the
instrument
so
that
air
can
circulate
through
it
freely.
When
operating
the
HP
8153A,
choose
a
location
that
provides
at
least
75mm
(3mches)
of
clearance
at
the
rear,
and
at
least
25mm
(linch)
of
clearance
at
each
side.
Failure
to
provide
adequate
air
clearance
may
result
in
excessive
internal
temperature,
reducing
instrument
reliability.
Input/Output
Signals
Caution
A
maximum
of
15V
can
be
applied
as
an
external
voltage
to
any
of
the
BNC
connectors.
HP-IB
Interface
You
can
connect
your
HP-IB
interface
into
a
star
network,
a
linear
network,
or
a
combination
star
and
linear
network.
The
limitations
imposed
on
this
network
are
as
follows;
The
total
cable
length
cannot
exceed
20
meters
The
maximum
cable
length
per
device
is
2
meters
No
more
than
15
devices
may
be
Interconnected
on
one
bus.
nstailation
A-S