EasyManuals Logo

HP 86140A Series User Manual

HP 86140A Series
304 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #154 background imageLoading...
Page #154 background image
4-48
Remote Operation
CALCulate Subsystem Commands
CALCulate:MARKer[1|2|3|4]:FUNCtion:BWIDth|BANDwidth[:STATe] OFF|ON|0|1
CALCulate:MARKer[1|2|3|4]:FUNCtion:BWIDth|BANDwidth[:STATe]?
Turns the bandwidth marker function
ON
or
OFF
for a particular marker. Only
one marker function is allowed to be on at a time for each individual marker.
Turning on the bandwidth function for a marker will turn off any other marker
function and turn on the Bandwidth function. For example, turning on the
Bandwidth function for a marker that has the Delta function
ON
, will turn
OFF
the Delta function and turn
ON
the Bandwidth function for the marker.
If the bandwidth function is turned
ON
for a marker that is
OFF
, the marker will
be turned
ON
, placed at the center wavelength, and then the bandwidth mark-
ers will measure the bandwidth relative to this marker.
CALCulate:MARKer[1|2|3|4]:FUNCtion:BWIDth|BANDwidth:X:CENTer?
Returns the absolute X-axis value from the center of the bandwidth marker
(mean of the left and right markers). The units returned are determined by
the CALCulate:MARKer:FUNCtion:BWIDth|BANDwidth:X:READout state. For
READout of FREQuency, the X value is returned in Hertz. For READout of
WAVelength, the X value is returned in meters.
This query generates a Settings conflict error if the bandwidth function is
OFF
for the specified marker.
CALCulate:MARKer[1|2|3|4]:FUNCtion:BWIDth|BANDwidth:X:LEFT?
Returns the absolute X-axis value of the left bandwidth marker. The units
returned are determined by the CALCulate:MARKer:FUNC-
tion:BWIDth|BANDwidth:X:READout state. For READout of FREQuency, the
X value is returned in Hertz. For READout of WAVelength, the X value is
returned in meters.
This query generates a Settings conflict error if the bandwidth function is
OFF
for the specified marker.
book.book Page 48 Monday, January 31, 2000 10:34 AM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 86140A Series and is the answer not in the manual?

HP 86140A Series Specifications

General IconGeneral
BrandHP
Model86140A Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals