4-55
Remote Operation
CALCulate Subsystem Commands
CALCulate:MARKer:SRANge:LOWer:FREQuency <param>
Sets the lower limit for the marker search range. Setting this value when
CALCulate:MARKer:SRANge:STATe is
OFF
will automatically turn
CALCulate:MARKer:SRANge:STATe
ON
. The range used for the marker search
range is the same range used for the total power calculation, the trace mean
range, and the wavelength sweep range. Changing the range with this com-
mand will change all four ranges. Sending the command when the instrument
is in a zero span will generate a Settings conflict error. The default units for
the parameter is in Hertz.
CALCulate:MARKer:SRANge:LOWer:TIME <param>
Sets the lower limit for the marker search range. Setting this value when
CALCulate:MARKer:SRANge:STATe is
OFF
will automatically turn
CALCulate:MARKer:SRANge:STATe
ON
. The range used for the marker search
range is the same range used for the total power calculation, the trace mean
range, and the wavelength sweep range. Changing the range with this com-
mand will change all four ranges. Sending this command while span is not set
to zero will results in a Settings conflict error. Default units for the parame-
ter is in seconds.
CALCulate:MARKer:SRANge:LOWer[:WAVelength] <param>
Sets the lower limit for the marker search range. Setting this value when
CALCulate:MARKer:SRANge:STATe is
OFF
will automatically turn
CALCulate:MARKer:SRANge:STATe
ON
. The range used for the marker search
range is the same range used for the total power calculation, the trace mean
range, and the wavelength sweep range. Changing the range with this com-
mand will change all four ranges. Sending the command when the instrument
is in a zero span will generate a Settings conflict error. Default units for the
parameter is in meters; frequency units are allowed.
book.book Page 55 Monday, January 31, 2000 10:34 AM