6-18
Reference
Cleaning Connections for Accurate Measurements
Cleaning a non-lensed connector
CAUTION
Do not use any type of foam swab to clean optical fiber ends. Foam swabs can
leave filmy deposits on fiber ends that can degrade performance.
1
Apply pure isopropyl alcohol to a clean lint-free cotton swab or lens paper.
2
Clean the ferrules and other parts of the connector while avoiding the end of
the fiber.
3
Apply isopropyl alcohol to a new clean lint-free cotton swab or lens paper.
4
Clean the fiber end with the swab or lens paper.
Do
not
scrub during this initial cleaning because grit can be caught in the
swab and become a gouging element.
Table 6-3. Cleaning Accessories
Item HP Part Number
Isopropyl alcohol 8500-5344
Cotton swabs 8520-0023
Small foam swabs 9300-1223
Compressed dust remover (non-residue) 8500-5262
Table 6-4. Dust Caps Available for Lightwave Instruments
Item HP Part Number
Laser shutter cap 08145-64521
FC/PC dust cap 08154-44102
Biconic dust cap 08154-44105
DIN dust cap 5040-9364
HMS10/HP dust cap 5040-9361
ST dust cap 5040-9366
book.book Page 18 Monday, January 31, 2000 10:34 AM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 86140A Series and is the answer not in the manual?

HP 86140A Series Specifications

General IconGeneral
BrandHP
Model86140A Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals