These commands call reflection standard classes during a calibration
sequence. If only one standard is in the class, it is measured. If there
is more than one, the standard being used must be selected with
STAN<AIBICIDIEIFIG>.
If there is only one standard in the class, these
commands are OPC-compatible.
CLASSllA
SllA:
Sll
l-port, opens.
CLASSllB
SllB:
Sll
l-port, shorts.
CLASSllC
SllC:
Sll
l-port, loads.
CLASS22A
S22A:
S22
l-port, opens.
CLASS22B
S22B:
S22
l-port, shorts.
CLASSBBC
S22C:
S22
l-port, loads.
CLEA<I>
Clears the indicated save/recall registers.
OPC-compatible. Range: 1 to 5.
CLEAL
Clears the limit line list. Should be
preceded by EDITLIML.
CLEARALL
Clears all the save/recall registers.
OPC-compatible.
CLEABIT[D]
Clears the specified bit on the GPIO.
Range: 0 to 7.
CLEABEG<I>
Clears save/recall registers 01 through
31.
CLEAREGOl
through
CLEAREG05
are the same as CLEAl through
CLEA5.
OPC-compatible. Range: 01 to 31.
CLEASEQ<I>
Clears the indicated sequence from the
internal registers. Range: 1 to 6.
CLEL
Clears the currently selected list. This
could be a frequency list, power loss list,
or limit test list.
CLES
Clears the status register, the event-status
registers, and the enable registers.
CLS
COAD
Same as CLES.
Selects coaxial electrical delay. See also
WAVD.
COAX
Selects coaxial offsets instead of
waveguide while defining a standard
during a cal kit modification.
l-108
HP-IB
Progr
amming and Command Reference Guide