Learn String and Calibration-Kit String
The learn string is a summary of the instrument state. It includes all the
front-panel settings, the limit-test tables, and the list-frequency table for
the current instrument state. It does not include calibration data or the
information stored in the save/recall registers.
The learn string can be output to a controller with the
OI-ITPLEHS
j
command, which commands the analyzer to start transmitting the binary
string. The string has a fixed length for a given firmware revision. The
array has the same header as in FORM 1. Refer to “Example 5: Using
the Learn String,” located in HP BASIC Programming Examples
Guide
The calibration kit includes a set of key characteristics of the calibration
standards used to determine the calibration accuracy. There are default
kits for several different connector types. There is also space for a
user-defined calibration kit. The command OUTPCALK outputs the
currently active calibration kit as a binary string in FORM 1. As with
the learn string, the calibration-kit string has a fixed length for a given
firmware revision.
1-42 HP-IB
Programm
‘ng and Command Reference Guide