EasyManua.ls Logo

HP 8753E - Calibration

HP 8753E
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration
Measurement calibration over HP-IB follows the same command
sequence as a calibration from the front-panel. For detailed information,
refer to “Optimizing Measurement Results” in the
I-LP
8753E
Network
Analum
User’s Guide
1. Start by selecting a calibration kit, such as 50 ohm type-N
(I:HLt:::t.158;
).
2. Select a calibration type, such as
Sll
l-port (CAL IS 1 1 1;
).
3. Call each class used by the calibration type, such as
F
111
R
i,J
A
R
D
:
0
P
E
t.4
(C:LHSS
11
H
j
)
During a 2-port calibration, the
reflection, transmission, and isolation subsequences must be opened
before the classes in the subsequence are called, and then closed at
the end of each subsequence.
4. If a class has more than one standard in it, select a standard from the
menu presented
(STAHA
to
STAHG).
5. If, during a calibration, two standards are measured to satisfy one
class, the class must be closed with
DOt4E
j
.
6. Declare the calibration done, such as with
D
l1lt.l
E
1
-
F’
0
R
T
1:
H
L
(SHV
1
j
over HP-IB).
The
STANA
to STANG commands will hold off the HP-IB until
completion because they trigger a sweep. If a class has only one
standard in it, which means that it will trigger a sweep when called, the
class command will also hold off the HI-IB.
Note
Since different cal kits can have a different number of
standards in a given class, any automated calibration
sequence is valid only for a specific cal kit.
1-54 HP-IB Programming and Command Reference Guide

Other manuals for HP 8753E

Related product manuals