CORI<ONIOFF>
Turns interpolative error correction ON
and OFF’.
CORR<ONIOFF>
COUC<ONIOFF>
Turns error correction ON and OFF.
Couples and uncouples the stimulus
between the channels.
COUP<ONIOFF>
Couple the power when coupled channels
is turned OFF, COUCOFF.
CSWI<ONIOF’F>
Cm=Q[D]
CWTIME
Selects the CW time sweep type.
DBXUPCHB
DBXUPCH3
D4XUPCH2
Selects test set continuous switching (ON)
or test set hold (OFF) when there is a
2-port calibration active. Continuous
switching is allowed only when the power
ranges on both attenuator ports are set
the same. When continuous switching
is ON, the analyzer measures all four
S-parameters each time before displaying
the data for a full 2-port cal measurement.
In test set hold mode, the analyzer
measures all four S-parameters once and
then measures the desired parameter
continuously. This is known as a fast
2-port cal measurement and it is less
accurate than a full 2-port calibrated
measurement.
Sets the CW frequency for power sweep
and CW frequency modes. While the list
frequency table segment is being edited, it
sets the center frequency of the current
segment. Range: stimulus range
Sets ups a two-graticule display with
channel 2 on top.
Sets ups a two-graticule display with
channel 3 on top.
Sets ups a four-graticule display with
channel 2 in the upper right quadrant of
the display.
l-l 10 HP-IB
Programmi
ng and Command Reference Guide