Chapter 4, Status Reporting
Status Reporting
Table 5 Calibration Status Register Group Bit Assignments
Bit
Number
Is Condition
Register
Implemented?
Condition/Event Comment
15 Not Used (Always 0) Defined by SCPI Version 1994.0
14 Unused in Test Set
13 Unused in Test Set
12 Unused in Test Set
11 Unused in Test Set
10 Unused in Test Set
9 Unused in Test Set
8 NO CDMA Baseband Physical Channel
Calibration Error
7 Unused in Test Set
6 NO Channel Power Calibration Failed Make sure cables are connected
between DUPLEX OUT or RF
Out Only (PCS Interface) and RF
IN/OUT connectors.
5 NO Digital Power Zeroing Failed CDMA Average Power measure-
ment zero attempt failed.
4 NO TX Power Auto-Zero Failed
3 YES Voltmeter Self-Calibration Failed
2 YES Counter Self-Calibration Failed
1 YES Sampler Self-Calibration Failed
0 YES Spectrum Analyzer Self-Calibration
Failed