82
S:\HP8924C\USRGUIDE\BOOK\chapters\opoverv.fb
Chapter 3, Operating Overview
Triggering Analog Measurements In Remote Mode (HP-IB Operation)
Triggering Analog Measurements In Remote Mode (HP-IB Operation)
1. Select the desired trigger mode.
Retriggering
Repetitive - Once a measurement cycle has completed
the Test Set is automat-
ically re-triggered and another measurement cycle begins.
When the trigger mode is set to repetitive retriggering, consecutive queries of the
same measurement return new measured values.
Single - Once a measurement cycle has completed the Test Set requires an
HP-IB trigger command be received to begin a new measurement cycle.
When the trigger mode is set to single retriggering, consecutive queries of the
same measurement (with no intervening trigger) will return the same value. Mea-
surements that rely on external signals or hardware-generated events (such as Traf-
fic Rho) must be re-armed with a new trigger before another measurement can be
made.
Settling
Full - Appropriate time delays introduced into measurement cycle to allow inter-
nal or external signal transients to settle.
Fast - No time delay introduced into measurement cycle to allow internal or ex-
ternal signal transients to settle.
The programmer must account for transient settling before issuing a trigger com-
mand.
There will still be delays introduced by the couplings between autotuning and
autoranging. If the programmer wishes to remove these delays as well, all autor-
anging and autotuning functions must be turned OFF and the program must explic-
itly set the ranging amplifiers and the frequency tuning. Delays introduced by the
measurement processes themselves cannot be eliminated.
HP-IB Syntax|
"TRIG:MODE:RETR REP" !selects Repetitive retriggering mode.
"TRIG:MODE:RETR SING" !selects Single retriggering mode.
TRIG:MODE:SETT FULL" !selects Full settling mode.
"TRIG:MODE:SETT FAST" !selects Fast settling mode.

Table of Contents