EasyManua.ls Logo

HP 8924C - Increasing Measurement Throughput

HP 8924C
350 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
86
S:\HP8924C\USRGUIDE\BOOK\chapters\opoverv.fb
Chapter 3, Operating Overview
Increasing Measurement Throughput
Increasing Measurement Throughput
This section contains a number of techniques for increasing measurement
throughput. These techniques include:
"Performing Transmitter and Receiver testing concurrently" on page 86
"Minimizing bus configuration time" on page 90
"Reducing delays caused by screen changes" on page 93
"Reducing delays caused by unused measurements" on page 92
"Reducing measurement setup time" on page 93
Performing Transmitter and Receiver testing concurrently
During an FER test, the Test Set is capable of executing HP-IB commands and
returning measurement results from other measurement functions, such as Rho.
Programming Example
The following example assumes a Service Option 2 call is Connected. The
program begins an FER measurement and allows an operator to query rho while
the FER frame count continues.