Chapter , Error Messages
The ROM Disk utility RAM_USAGE will display the total amount of non-volatile RAM
installed in the Test Set, the RAM Disk allocation, the Save/Recall register allocation and
the amount of non-volatile RAM available to IBASIC.
HP-IB Error: Unknown Save/Recall error.
This error can occur when you are trying to SAVE the instrument state to a mass storage
device with a LIF formatted media. The default file system is DOS. Refer to chapter 5,
"Memory Cards/Mass Storage" for more information.
HP-IB Error: HP-IB Units cause invalid conversion of attr.
This error is generated when trying to change Attribute Units and one of the Data Function
values is set to zero. If this error is encountered the programmer must change the Data
Function settings to values that can be converted to the new units_of_measure. Refer to
"To Specify Units-of-Measure for HP-IB Data Transfer" on page 56 for more details.