EasyManua.ls Logo

HP 8924C - Page 34

HP 8924C
350 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
34
S:\HP8924C\USRGUIDE\BOOK\chapters\getstart.fb
Chapter 1, Getting Started
Obtaining Measurement Results
To Make a Simple Measurement
1. Access the screen containing the desired measurement by using the DISP command.
2. Use the MEAS form of the syntax for that measurement to place the measured value
into the Test Set’s output buffer.
3. Enter the value into the correct variable type within the program’s context (refer to the
”Number Measurement Syntax” in HP-IB Command Syntax chapter of the HP 8924C
Condensed Programming Reference Guide for the proper variable type).
NOTE: Whenever a numeric value is queried, the returned value is always in HP-IB Units. Refer to
"To Specify Units-of-Measure for HP-IB Data Transfer" on page 56.
The following example program illustrates how to make settings and then take a
reading from the Test Set. This setup takes a reading from the analog spectrum
analyzer’s marker after tuning it to the RF generator output frequency.
10 Addr=714
20 OUTPUT Addr;”*RST”! Preset to known state
25 OUPTUT Addr;”CONF:OFR 0” !Set the RF frequency offset to 0 MHz.
30 OUTPUT Addr;”TRIG:MODE:RETR SING”! Sets single trigger
40 OUTPUT Addr;”DISP RFG”! Selects the RF GENERATOR screen
50 OUTPUT Addr;”AFG1:FM:STAT OFF”! Turns FM OFF
60 OUTPUT Addr;”RFG:AMPL -66 DBM”! Sets RF Gen amplitude to -66 dBm
70 OUTPUT Addr;”RFG:FREQ 500 MHZ”! Sets RF Gen frequency to 500 MHz
80 OUTPUT Addr;”RFG:AMPL:STAT ON”! Turns RF Gen output ON
90 OUTPUT Addr;”DISP SAN”! Selects SPECTRUM ANALYZER screen
100 OUTPUT Addr;”SAN:CRF 500 MHZ”! Center Frequency 500 MHz
110 ! -------------------MEASUREMENT SEQUENCE--------------------
120 OUTPUT Addr;”TRIG”! Triggers reading
130 OUTPUT Addr;”MEAS:SAN:MARK:LEV?”! Query of Spec Anl marker level
140 ENTER Addr;Lvl! Places measured value in variable Lvl
150 DISP Lvl! Displays value of Lvl
160 END
The above example is very simple and is designed to demonstrate the fundamental
procedure for obtaining a measurement result. Many other factors must be
considered when designing a measurement procedure, such as instrument settings,
signal routing, settling time, filtering, triggering and measurement speed.

Table of Contents

Related product manuals