64
Over-temperature protection
CONTROL OF PROTECTION:
Stop
Process security
Go back
Stop: cutoff of heating, pump
and refrigeration machine
Process security:
refrigeration machine continues.
CHARACTERISTICS:
OT limit: heating
OT Mode
Display all OT values
Return to main menu
CHARACTERISTICS:
OT limit: heating
OT Mode
Display all OT values
Return to main menu
Code for OT input
77
Please enter code
Actual setting: 0000
Max. setting 65535
Min. setting 0000
New value xxxx
Notice:
Please consider the statutory and
internal regulatons when setting
the OT limit.
OT Set-point Circulation
Actual setting: 35.00
Temp. max. 200.00
Act. temp. 22.40
Temp. min -40.00
New setting xxxx
Protection overtemperature
Actual OT values
Sensor heater 1 19.0°C
Tripping points
Setting OT ht.1 35.0°C
OT Mode Stop
Continue: Press Encoder
OK
ESC
KEYB
ESC
OK
KEYB
OK
OK
OT Mode
OT limit: heating

Table of Contents