SR05-D1A3-PV manual v1801 77/83
Table 9.11.3 Modbus registers 0x1000 to 0x1052
MODBUS REGISTERS 0x1000 0x1052
REGISTER
ADDRESS
(HEX)
PARAMETER DESCRIPTION OF CONTENT TYPE
OF
(R/W)
FORMAT
OF DATA
0x1000 Modbus address Sensor address in Modbus
network, default = 1
R/W U16
0x1001
Serial communication
settings
Sets the serial
communication, default = 5
R/W
U16
0x1002 +
0x1003
Irradiance
signal in x 0.01
W/m² R S32
0x1004 +
0x1005
Factory use only
0x1006 Sensor body
temperature
In x 0.01
°
C R S16
0x1007
Sensor electrical
resistance
In x 0.1
R
U16
0x1008 Scaling factor irradiance Default = 100 R U16
0x1009 Scaling factor
temperature
Default = 100 R U16
0x100A +
0x100B
Sensor voltage output
In x 10
-9
V
R
S32
0x100C to
0x101F
Factory use only
0x1020 to
0x1023
Sensor model
Part one of sensor description
R
String
0x1024 to
0x1027
Sensor model Part two of sensor description R String