Troubleshooting
34
Troubleshooting
Problem:
USB cable connected, but the battery is not being charged
Reason: USB port of laptop does not provide enough power
Solution: Connect the meter to the battery charger or a USB port which is able to supply at
least 500 mA.
Reason: USB cable broken
Solution: Replace cable
Problem:
The meter can not be switched on
Reason: Battery is empty.
Solution: Connect the meter to a battery charger or an USB port.
Reason: Other problems.
Solution: Press and hold the on/off key for at least 5 seconds. The meter will perform a
"Low-Level-Reset" and should start afterwards.
Problem:
The meter powers off itself after a couple of seconds.
Reason: Battery empty.
Solution: Connect the meter to a battery charger or an USB port.
Problem:
pH channel can not be selected (only Version BP)
Reason: pH channel is disabled in basic settings menu.
Solution: enable pH channel in basic settings menu.
Problem:
Pressure value does not tare / zero balancing does not work (only versions BO and BP)
Reason: There is still pressure applied to the pressure transducer.
Solution: Disconnect pressure hose directly at meter.