EasyManuals Logo

IBP HDM97BQ User Manual

Default Icon
50 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
Troubleshooting
34
Troubleshooting
Problem:
USB cable connected, but the battery is not being charged
Reason: USB port of laptop does not provide enough power
Solution: Connect the meter to the battery charger or a USB port which is able to supply at
least 500 mA.
Reason: USB cable broken
Solution: Replace cable
Problem:
The meter can not be switched on
Reason: Battery is empty.
Solution: Connect the meter to a battery charger or an USB port.
Reason: Other problems.
Solution: Press and hold the on/off key for at least 5 seconds. The meter will perform a
"Low-Level-Reset" and should start afterwards.
Problem:
The meter powers off itself after a couple of seconds.
Reason: Battery empty.
Solution: Connect the meter to a battery charger or an USB port.
Problem:
pH channel can not be selected (only Version BP)
Reason: pH channel is disabled in basic settings menu.
Solution: enable pH channel in basic settings menu.
Problem:
Pressure value does not tare / zero balancing does not work (only versions BO and BP)
Reason: There is still pressure applied to the pressure transducer.
Solution: Disconnect pressure hose directly at meter.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the IBP HDM97BQ and is the answer not in the manual?

IBP HDM97BQ Specifications

General IconGeneral
BrandIBP
ModelHDM97BQ
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals