35 Interface
Technical data
Conductivity
0 to 30,0 mS/cm, 4-Pole probe,
Resolution and accuracy:
Reference temperature 25° C Temperature compensation is 2,070 % per Kelvin
Adjustable from 0.00 to 4,00 %/Kelvin
Or selectable from dialysis machine manufacturer preset values.
Temperature (only versions BN, BO, BP and BQ)
Range 0 … 100 °C,
Resolution 0,01 °C
Accuracy 0,1°C
Pressure all versions
Units mmHg, kPa, mbar, PSI
Versions BO, BP and BQ
Range -700 … 1900 mmHg, 2.5 Bar
Resolution 0,1 mmHg
Accuracy 0 … 300 mmHg 1 mmHg
Otherwise 2 mmHg
Fluid resistance yes
Version BM and all other versions with option H)
Range -700 … 1900 mmHg, 2.5 Bar
Resolution 0.01 mmHg
Accuracy -299.9 … 300 mmHg 0.5 mmHg
Otherwise 1 mmHg
Fluid resistance no