EasyManua.ls Logo

Icom IC-R8600 - To Remove a Sharp Spike When Setting a Low Sampling Rate Value

Icom IC-R8600
10 pages
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5
A7438-2EX © 2017 Icom Inc.
To remove a sharp spike when setting a low sampling rate value
If the sampling rate value is too low, a sharp spike (a DC) is displayed at the LO frequency position.
The DC is a constant offset in voltage, the Left and Right (I and Q) ADCs produce, due to error.
To remove the sharp spike:
Set a higher sampling rate value.
or
Adjust the IQ balance. (Option [F7] Calibration Settings DC Removal Calibration for RX)
To adjust the IQ balance
1. Set RF Gain to “MIN” on the IC-R8600 I/Q Settings screen. (p. 3)
Only the sharp spike is displayed.
2. Open “DC Removal Calibration for RX.
3. Click the displayed mode to set “IIR-Highpass (Auto).
The software automatically adjusts the IQ balance.
4. Set Mode to “Constant (On).
L If this setting remains “IIR-Highpass (Auto),” the software removes signals near the LO frequency.
5. Adjust RF Gain on the IC-R8600 I/Q Settings screen.
NOTE: When you change the sampling rate value, adjust the IQ balance again.
a sharp spike
(a DC)

Other manuals for Icom IC-R8600

Related product manuals