COMMAND REFERENCE
Command Reference
7/01 LDC-3700 Series 91
CHAPTER 4
The TEC:CAL:ITE command is used to enter the TEC’s current set point, measurement, and limit
calibration mode.
Parameters None.
Notes This is a service related command. When this command is issued, the front panel becomes
disabled, and the appropriate adjustments are made to the operating modes. This mode should not
be entered unless the user has the proper measurement equipment in place. See Chapter 6.
After this command is issued, the LDC−3700B Series Laser Diode Controller will automatically
change to ITE mode, turn the TEC output on, determine its zero current offset, set the ITE limit to
4.0 amps, and drive the output to 1.0 amps. This procedure is outlined in Chapter 6.
The TEC:CAL:ITE? query (or bit 11 of the TEC status condition register) may be used to determine
if the LDC−3700B Series Laser Diode Controller is ready for the user to enter a value.
Examples "Tec:CAL:ITE" −action: the LDC−3700B Series Laser Diode Controller enters calibration mode for
current.
"Tec:Cal:Ite" −action: the LDC−3700B Series Laser Diode Controller enters calibration mode for
current.
TEC:CAL:ITE
COMMON
DEVICE DEPENDENT
FRONT PANEL