COMMAND REFERENCE
Command Reference
60 LDC-3700 Series
CHAPTER 4
Action Enables bits in the standard event status enable register.
Response The value must be between 0 and 255.
Bit Reference
Notes Bit 5 of the status byte register is set if any enabled conditions are true.
Setting bit 0 allows you to generate service requests from overlapped commands as previous oper−
ations complete. This may be useful for ensuring that an operation is complete before starting a
measurement.
*ESE <nrf value>
Event Status Enable
COMMON
DEVICE DEPENDENT
FRONT PANEL
0− Operation Complete 4− Execution Error
1− N/A 5− Command Error
2− Query Error 6− User Request
3− Device Dependent Error 7− Power On
76543210
76543210
&
&
&
&
&
&
&
&
Logical OR
Standard Event Status Register
Standard Event Status Enable Register
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com