COMMAND REFERENCE
Command Reference
68 LDC-3700 Series
CHAPTER 4
The LASer:DISplay:LDI command sets the laser display to show the constant current measurement.
Parameters None.
Notes The actual LASER I display is turned off automatically when another LASER DISPLAY selection is
enabled.
Examples "LAS:DIS:LDI" −action: enables the LASER display for current values.
The LASer:DISplay:LDI? query returns the status of the (LAS DISPLAY) I switch.
Notes In local operation, the status of the I switch is indicated by the enunciator next to the LASER display
(mA will be illuminated when the display is in I mode).
Examples "LAS:DIS:LDI?" −response: 0, means that the (LASER DISPLAY) I switch is not currently active,
laser current is not displayed.
"Las:dis:ldI?" −response: 1, means that the (LASER DISPLAY) I switch is currently active, laser
current may be displayed.
The LASer:DISplay:LDV command sets the laser display to show the laser forward voltage
measurement.
Parameters None.
Notes The actual LASER V display is turned off automatically when another LASER DISPLAY selection is
enabled.
Examples "LAS:DIS:LDV" −action: enables the LASER display for forward voltage values in Volts.
The LASer:DISplay:LDV? query returns the status of the (LAS DISPLAY) V switch.
Notes In local operation, the status of the V switch is indicated by the enunciator next to the LASER
display (V will be illuminated when the display is in V mode).
Examples "LAS:DIS:LDV?" −response: 0, means that the (LASER DISPLAY) V switch is not currently active,
laser voltage is not displayed.
"LAS: DIS:LDV?" −response: 1, means that the (LASER DISPLAY) V switch is currently active,
laser voltage may be displayed.
LASer:DISplay: LDI
I
COMMON
DEVICE DEPENDENT
FRONT PANEL
LASer:DISplay: LDI?
I?
COMMON
DEVICE DEPENDENT
FRONT PANEL
LASer:DISplay: V
LDV
COMMON
DEVICE DEPENDENT
FRONT PANEL
LASer:DISplay: V?
LDV?
COMMON
DEVICE DEPENDENT
FRONT PANEL