54
ETSI EN 301 489-1 and ETSI EN 301 489-27
GUIDELINES AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC
EMMISSIONS OF THE GUARDIO CHARGER PURSUANT TO:
ETSI EN 301 489-1 V2.2.3 ElectroMagnetic Compatibility (EMC) standard for radio
equipment and services; Part 1: Common technical requirements; Harmonised
Standard for ElectroMagnetic Compatibility
ETSI EN 301 489-27 ElectroMagnetic Compatibility (EMC) standard for radio
equipment and services; Part 27: Specific conditions for Ultra Low Power Active
Medical Implants (ULP-AMI) and related peripheral devices (ULP-AMI-P) operating
in the 402 MHz to 405 MHz bands; Harmonised Standard covering the essential
requirements of article 3.1(b) of Directive 2014/53/EU
The Guardio Charger, part of the OPTIMIZER Integra CCM-D System, is intended for use
in an electromagnetic environment as specified below. The customer or user of the
Guardio Charger must ensure that it is used within the specified environment.
There may be potential difficulties in ensuring electromagnetic compatibility in other
environments, due to conducted as well as radiated disturbances.
Emissions Test
Basic Standard
Compliance
Electromagnetic
environment
guidelines
Radiated Emissions
EN 55032
N/A covered
by relevant
radio
standards
The Guardio Charger
must emit electromagnetic
energy in order to perform
its intended function.
Nearby electronic
equipment may be
affected.
Conducted Emissions
EN 55032
Pass
Harmonic Current
Emissions
IEC 61000-3-2
Pass
Voltage Fluctuations
IEC 6100-3-3
Pass