tirb45e1  (2019-02) CDGxxxDxx.cp    45 
 
Percentage of High Trip Source: Percentage of Piezo Sensor Value (I 0x6010, 
SI 0x11) in [%]. 
 
 
Low Trip Point Limit: Low limit to trigger trip point condition if Input Value (I 0x900E, 
SI 0x01) is below this limit. 
 
 
Object index of Low Trip Point Value source. Index of Low Trip Point Limit 
(I 0x800E, SI 0x14) 
 
Bit 16…31: Index 
Bit 08…15: Subindex 
Bit 00…07: reserved 
It is only possible to reference to "Low Trip Point Limit" (I 0x800E, SI 0x14) or 
"Input Piezo Module 2" "Sensor Value" (I 0x6010, SI 0x11). 
 
 
Percentage of Value referenced by Low Trip Source Index in [%]. If Low Trip 
Source Index is I 0x800E, SI 0x14 this parameter is not used. 
 
 
High Trip Hysteresis: Hysteresis value for High Trip Point. 
 
 
Low Trip Hysteresis: Hysteresis value for Low Trip Point. 
 
 
Object index of active source of (I 0x900E SI 0x01) Input Value. 
 
Bit 16…31 
Bit 08…15 
Index 
Subindex 
 
 
 
Calculating the High Trip Point and Low Trip Point → chapter 5.5.2. 
 
Index  SI  DataType  NV  Access  PM  Name 
  0x13  REAL  x  RW    Percentage High Trip 
  0x16  REAL  x  RW    Percentage Low Trip 
 
 
0 
1 
Disable 
Enable 
 
 
 
5.5.4  Configuration Trip