48 tirb45e1 (2019-02) CDGxxxDxx.cp
Index
SI
DataType
NV
Access
PM
Name
9001
0x01
UINT
RO
Sensor Warning
0x02
UINT
RO
Sensor Error
Sensor Warnings
Bit 0
Bit 1…13
Bit 14
Bit 15
Not at temperature
0
Low Warning Exception
High Warning Exception
Sensor Errors
Bit 0
Bit 1
Bit 2
Bit 3…13
Bit 14
Bit 15
Diaphragm Failure
Electronics Failure
Electronics Over Temperature
0
Low Error Exception
High Error Exception
Index
SI
DataType
NV
Access
PM
Name
9010
0x01
REAL
x
RO
Accumulated Zero Offset
0x02 REAL x RO Highest Informational
Measurement Value
0x03 REAL x RO Highest Precision
Measurement Value
0x04 REAL x RO Lowest Precision
Measurement Value
Accumulated Zero Offset: An amount added prior to Gain to derive Value. Result of
Zero Adjust Command (accumulated value).
Highest Informational Measurement Value: Highest value that the gauge can
measure without a specified accuracy.
Highest Precision Measurement Value: Highest valid value at which the gauge is
specified with an accuracy value. Above that value, the parameter Reading Valid is
set to invalid.
Lowest Precision Measurement Value: Lowest valid value at which the gauge is
specified with an accuracy value. Below that value, the parameter Reading Valid is
set to invalid.
Index
SI
DataType
NV
Access
PM
Name
9012
0x01
UINT
RO
Sensor Warning
0x02
UINT
RO
Sensor Error
Sensor Warnings
Bit 0…15
0
Sensor Errors
Bit 0
Bit 0…15
Sensor Failure Piezo Sensor
0
5.6.2 Information Capacitance
Diaphragm
Subindex 0x01
Subindex 0x02
5.6.3 Information Common
Piezo
Subindex 0x01
Subindex 0x02
Subindex 0x03
Subindex 0x04
5.6.4 Information Piezo
Subindex 0x01
Subindex 0x02

Table of Contents