tirb45e1  (2019-02) CDGxxxDxx.cp    57 
 
Execution of this command will start a Zero Adjust operation. 
  The zeroing function is locked in order for operating errors to be 
prevented, therefore see the operation instructions of the vacuum 
gauges, refer to the appropriate documents (→  [1] … [8]). 
 
 
 
0: Zero adjust with no offset 
Index of the Sub Sensor Instance (1) 
Offset value (Data format: REAL, always 0) 
 
 
Status (supported values) 
0 
1 
2 
 
3 
Last command completed, no errors, no reply available 
Last command completed, no errors, reply available 
Last command completed, errors present, no reply 
available 
Last command completed, errors present, reply available 
 
 
Byte 2  0: Zeroing successful 
1: Zeroing failed; out-of-range 
2: Zeroing failed; cumulative out-of-range 
3: Zeroing failed: measurement invalid 
254: No previous Zero Adjust command issued 
 
 
 
Execution of this command will start a Full Scale Adjust operation. 
 
 
 
2: Index of the Sub Sensor Instance (always 2) 
0: Full Scale value (Data format: REAL, always 0) 
 
 
Status (supported values) 
0 
1 
2 
 
3 
Last command completed, no errors, no reply available 
Last command completed, no errors, reply available 
Last command completed, errors present, no reply 
available 
Last command completed, errors present, reply available 
 
5.8.2  Command Zero Adjust