46
tina60e1-a (2011-04)
Data Rate Setting
By means of the <RATE> switch, the data rate can
be set to 125 (<1>), 250 (<2>) or 500 kBaud (<5>)
(default 500 kBaud).
If the switch is in the <P> position, the data rate is
programmable via the DeviceNet (→ [4]).
Status LED
Two LEDs on the gauge inform on the gauge status and the
current DeviceNet status.
<STATUS MOD> (gauge status):
LED Meaning
off No supply
blinking green-red Selftest
lit solid green Normal operation
lit solid red Non recoverable error
blinking red Recoverable error (e.g. missing
DeviceNet power supply)