tina60e1-a (2011-04) 47
<STATUS NET> (network status):
LED Meaning
off Gauge not online:
Selftest not yet concluded
No supply, "STATUS MOD"
blinking green Gauge online but no communication:
Selftest concluded but no communica-
tion to other nodes established
Gauge not assigned to any master
lit solid green Gauge online; necessary connections
established
blinking red One or several input / output connections in
"time out" status
lit solid red Communication error. The gauge has de-
tected an error that impedes communica-
tion via the network (e.g. two identical node
addresses (MAC IC) or "Bus-off")