Section 8: System Specifications
32 of 34
NOTE - U
T
is the AC mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
The InMode System with Morpheus8 Body Handpiece is intended for use in the electromagnetic
environment specified below. The customer or the user of the InMode System with Morpheus8
Body Handpiece should assure that it is used in such an environment.
IMMUNITY
Test
IEC 60601
Test
Level
Compliance
Level
Electromagnetic environment guidance
Conducted
RF IEC
61000-4-6
Radiated RF
IEC 61000-
4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2,5 GHz
[3] V
[3] V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the device, including cables, than the
recommended separation distance calculated
from the equation applicable to the
frequency of the transmitter.
Recommended separation distance


 󰇟󰇠

 󰇟󰇠

Where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a Field-strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess