Morpheus8 Body Applicator Operator
Manual
Section 8: System Specifications
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the
InMode System with Morpheus8 Body Handpiece is used exceeds the applicable RF
compliance level above, the InMode System with Morpheus8 Body Handpiece should be
observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the InMode System with
Morpheus8 Body Handpiece.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications
equipment and the InMode System with Morpheus8 Body Handpiece System
The InMode System with Morpheus8 Body Handpiece System is intended for use in an
electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the InMode System with Morpheus8 Body Handpiece can help
prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the InMode
System with Morpheus8 Body Handpiece as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output
power of transmitter
[W]
Separation distance according to frequency of transmitter [m]
150 kHz to 80 MHz
80 MHz to 800
MHz
800 MHz to 2,5
GHz
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.