EasyManuals Logo

Instron 3400 Series User Manual

Instron 3400 Series
126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #114 background imageLoading...
Page #114 background image
Chapter: Maintenance
114 M10-17313-EN
Lubrication procedures
Refer to “Lubricate ballnuts” on page 114 if your load frame model is 34SC-05,
34SC-1 or 34SC-2.
Refer to “Lubricate ballnuts and linear guide carriages” on page 116 if your load
frame model is 34SC-5.
Lubricate ballnuts
This procedure applies to models 34SC-05, 34SC-1, and 34SC-2, refer to Figure 23 on
page 115.
Table 10. Lubrication Requirements for Ballnuts
Recommended Lubricant Interval Instron
®
Part Number
Lithium based water resisting grease -
NLGI Class 2
(DIN51825 class K2k LS2)
Equivalents:
Lubriplate: Lubriplate 1200-2
Mobil: Mobilux® 2
BP: Energrease® LS2
Shell: Alvania® R2
1000
1
hours or 36
months of
operating time
2
105-1-1057
(small can of lubricant)
1. If the system is used more than 100 hours per month, inspect the lubricated area periodically and
increase lubrication frequency as required.
2. Operating time is defined as time of crosshead movement, not the length of time the system has been
powered on.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Instron 3400 Series and is the answer not in the manual?

Instron 3400 Series Specifications

General IconGeneral
BrandInstron
Model3400 Series
CategoryTest Equipment
LanguageEnglish