InstroTek
22
11. Device Configuration
The machine is capable of different loading histories, such as
the AASHTO TP 124 (I-FIT), ASTM D8044 (LSU), AASHTO T 283,
IDEAL-CT, and Marshall Stability, as well as other loading
histories. The machine can be programmed for different
loading histories using the information in this section.
1. Press on the main menu to enter the test.
2. To review the parameters for a standard, press
to see the parameters.
Loading History Parameters
Parameter
Description
Pre-load rate
Rate for applying the pre-load (seating load) (mm/min or
kN/sec).
Pre-load force
Target load for the pre-loading to stop
Pre-load pause
Time that the pre-load is held so the user can adjust the
specimen and record the pre-load
Pre-load release
Release of pre-load
(None, Timed, and Contact Load)
Rate
Loading rate for test
(mm/min or kN/sec)
Start load
Load level when data acquisition starts
Stop criterion
Criterion when machine stops loading
(Load, Time, or Percentage)