AA222 Instructions for Use US Page 27
3.6 Operating instructions Impedance
3.6.1 Calibration cavities
You can use the 0.2ml, 0.5ml, 2.0ml and 5ml for v daily validity checking of the probe calibration.
To perform a calibration check, select a protocol that measures a tympanogram.
Do not use an ear tip! Place the probe tip completely into the cavity. Perform the measurement. Check the
volume that was measured.
The permitted tolerance in the volume measurement is ±0.1ml for cavities up to 2ml and ±5% for larger
cavities. These tolerances are applicable for all probe tone frequencies.
We strongly recommend calibrating probe and contra phone at least once a year.
3.6.2 Handling and selection of ear tips
When using the AA222 probe and CIR contra phone, Sanibel
TM
ear tips must be used.
The Sanibel
TM
ear tips are single use only and should not be reused. Reuse of ear tips can lead to the
spread of infection from patient to patient.
The probe and CIR contra phone must be fitted with an ear tip of a suitable type and size before testing.
Your choice will depend on the size and shape of the ear canal and ear. Your choice may also depend on
personal preference and the way you perform your test.
When performing a quick impedance screening test you may choose an umbrella-shaped ear
tip. Umbrella ear tips seal the ear canal without the probe tip going into the ear canal itself.
Press the ear tip firmly against the ear canal so that a good seal is maintained throughout the
entirety of the test.
For more stable testing, we recommend using an extension cord with a mushroom-shaped
ear tip. Make sure that this ear tip inserts completely into the ear canal. Mushroom-shaped
ear tips allow you to test ‘hands free’ from the AA222. This reduces the chance of contact
noise disturbing the measurement.
To optimize the stability of the measurements, it is recommended to not hold the probe between the fingers
during the?testing. Acoustic reflex measurements, in particular, could be affected by probe movements.
CAUTION

Table of Contents

Other manuals for Interacoustics AA222

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Interacoustics AA222 and is the answer not in the manual?

Interacoustics AA222 Specifications

General IconGeneral
BrandInteracoustics
ModelAA222
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals