EasyManuals Logo

Interacoustics AA222 User Manual

Interacoustics AA222
93 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #34 background imageLoading...
Page #34 background image
AA222 Instructions for Use US Page 28
3.6.3 Probe status
The probe status is indicated by the color of the light on the control panel, the standard probe system and
the clinical probe system. Below, the colors and their meanings are explained:
Color
Control panel
Standard probe
Clinical probe
Status
Red
Right ear is selected. Probe is out of
the ear.
Blue
Left ear is selected. Probe is out of
the ear.
Green
Probe is in the ear and a seal is
maintained.
Yellow
Probe is in the ear and is blocked,
leaking or too noisy.
White
The probe has just been attached.
Probe status is unknown. If the probe
light stays white in any other
situation, the AA222 may need to be
switched off and back on again to
regain proper probe status.
Blinking
The AA222 is pausing and/or an
interaction is expected. The AA222
will, for example, remain blinking
green if the protocol has finished
testing and the probe is still in the
ear. Or the user may pause the
AA222 before inserting the probe,
resulting in blue or red blinking.
No light
AA222 is not monitoring the probe
status.

Table of Contents

Other manuals for Interacoustics AA222

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Interacoustics AA222 and is the answer not in the manual?

Interacoustics AA222 Specifications

General IconGeneral
BrandInteracoustics
ModelAA222
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals