AA222 Instructions for Use – US Page 68
5.1 Calibration Properties
Telephonics TDH39/DD45 with a static force of 4.5N
0.5N and/or EARtone 3A and/or CIR insert phone
Ipsilateral Earphone: is integrated in the probe system
Probe frequency transmitter and receiver and
pressure transducer is integrated in the probe system
Generally the instrument is made and calibrated to be
within and better than the tolerances required in the
Contralateral Reflex and
Audiometer Tone Levels:
3 dB for 250 to 4000Hz and 5 dB for 6000 to 8000Hz
Ipsilateral Reflex Tone
Levels:
5 dB for 500 to 2000Hz and +5/-10 dB for 3000 to
4000Hz
Pressure measurement :
Compliance measurement:
5% or 10 daPa, whichever is greater
5% or 0.1 ml, whichever is greater
Stimulus Presentation
Control:
Rise time = 20 ms
Fall time = 20 ms
A weighted SPL in Off = 31 dB
Impedance Calibration Properties
226 Hz 1%, 678 Hz 1%, 800 Hz 1%, 1000 Hz 1%
85 dB SPL 1.5 dB measured in an IEC 60318-5
acoustic coupler. The level is constant for all volumes
in the measurement range.
Reflex artifact level:
0.001 ml is the lowest detectable volume change
≥95 dB SPL (measured in the 711 coupler, 0.2 ml, 0.5
ml, 2.0 ml & 5.0 ml hard walled cavities).
characteristics:
(IEC60645-5 clause 5.1.6)
Initial latency = 35 ms (5 ms)
Rise time = 42 ms (5 ms)
Terminal latency = 23 ms (5 ms)
Fall time = 44 ms (5 ms)
Overshoot = max. 1%
Values between -600 to +400 daPa can be selected in
the setup.
-750 daPa and +550 daPa, 50 daPa
chances influence on the
impedance measurement
in the specified range
(97300 –
105300calibration
Pascal).
Admittance can vary inside: ± 4%
The pressure accuracy is: ±10 daPa or 10%,
whichever is greater.
The pressure sensor used, is a differential/gauge type, which means, it measure
the pressure difference and therefore not affected of the height above sea level.