AA222 Instructions for Use – US Page 73
5.2.2 Audiometry – Survey of reference and max hearing level tone audiometry
DD45 6ccm uses IEC60318-3 or NBS 9A coupler and RETSPL comes from PTB – DTU report 2009-2010.
Force 4.5N ±0.5N.
TDH39 6ccm uses IEC60318-3 or NBS 9A coupler and RETSPL comes from ANSI S3.6 2010 and ISO 389-
1 1998. Force 4.5N ±0.5N.
HDA280 6ccm uses IEC60318-3 or NBS 9A coupler and RETSPL comes from ANSI S3.6 2010 and PTB
2004. Force 5.0N ±0.5N.
HDA300 Artificial ear uses IEC60318-1 coupler with type 1 adaptor and RETSPL comes from PTB report
2012. Force 8.8N ±0.5N.
IP30 / EAR3A/EAR 5A 2ccm uses ANSI S3.7-1995 IEC60318-5 coupler (HA-2 with 5mm rigid Tube) and
RETSPL comes from ANSI S3.6 2010 and ISO 389-2 1994.
CIR 2ccm uses ANSI S3.7-1995 IEC60318-5 coupler HA2 and RETSPL uses the Insert value from comes
from ANSI S3.6 2010 and ISO 389-2 1994.
B71 / B81 uses ANSI S3.13 or IEC60318-6 2007 mechanical coupler and RETFL come from ANSI S3.6
2010 and ISO 389-3 1994. Force 5.4N ±0.5N.