AA222 Instructions for Use – US Page 48
3.7.1.10 SISI (short increment sensitivity index)
SISI is designed to test the ability to recognize a 1dB increase in intensity during a series of bursts of pure
tones presented 20 dB above the pure tone threshold for the test frequency. It can be used to differentiate
between cochlear and retrocochlear disorders, as a patient with a cochlear disorder will be able to perceive
the increments of 1 dB, whereas a patient with a retrocochlear disorder will not. 20 measures must be
obtained in order to have the SISI threshold shown at the given frequency.
Function Key
Description
A
Amplitude Modulation (0, 1(SISI), 2, 5).
B Start SISI test.
Pause SISI test.
C Stop SISI test.