AA222 Instructions for Use – US Page 86
(b)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
To ensure compliance with the EMC requirements as specified in IEC 60601-1-2, it is essential to use only
the following accessories:
ITEM
MANUFACTURER
MODEL
Power supply unit UE60
Interacoustics
UE60-240250SPA3
Clinical Probe
Interacoustics
Clinical probe system 1077/1078
Diagnostic probe
Interacoustics
Diagnostic probe system 1077/1078
DD45C Contra Headset DD45C contra
headset P3045
Interacoustics
DD45C
IP30 Insert Phone 10ohm single contra
Interacoustics
IP30C
DD45 Audiometric Headset P3045
Interacoustics
DD45
IP30 Insert Phone 10ohm set
Interacoustics
IP30
B71/B81 Bone conductor headset 10
Ohm(lead-free)
Interacoustics
B71 /B81
Conformance to the EMC requirements as specified in IEC 60601-1-2 is ensured if the cable types and cable
lengths are as specified below:
Description
Length
Screened/Unscreened
Mains Cable
2.0m
Unscreened
USB Cable
2.0m
Screened
Clinical Probe
2.0m
Unscreened
Diagnostic Probe
2.0m
Unscreened
DD45C contra headset P3045
2.0m
Screened
IP30 Insert Phone 10ohm single contra
2.0m
Screened
DD45 Audiometric Headset P3045
2.0m
Screened
IP30 Insert Phone 10ohm set
2.0m
Screened