8Troubleshooting
8.1Troubleshooting
EffectPossiblecauseVericationCorrectiveaction
Floorlocknotengaged.Greenplungeronoorlock
postisvisible.
Activateoorlock-green
plungershouldNOTbevisible.
Insufcientstaticfriction
betweenoorlockfoot
bumperandoor.
Floorsurfaceisslippery
and/orforeignmembrane
betweenbumperandoor.
Cleanthesurroundingoor
andremoveloosematerial
underthebed.
Beddoesnotstayinplace
Floorlocknotfunctioning.Floorlockisstuckinone
position.
ContactInvacareContinuing
CareServiceDepartment
1-800-668-2337.
Bedisnotlevel
Hi/Loactuatorsareunevenly
extended
Beddeckisnotparallelwith
oor
Raisebedtohighestposition
or
Lowerbedtolowestposition
Actuatornotworking.SystemFault
NeedsReset
FlashingLEDsonACP ,Bed
functionwillnotoperate.The
controlboxmaybeepwhen
pendantfunctionispressed.
ToReset-holdthependant
bedupandbeddownbuttons
for5+seconds.Thebedwill
beepandstopbeepingwhen
resetiscomplete.Thenpress
thebedupbutton.Ifthe
bedbeeps,theactuatorsare
outofsync.Continuerising
tothehighestpositionto
synchronizetheactuators.
56