EffectPossiblecauseVericationCorrectiveaction
Actuatornotworking.Faultyactuator.Disconnectpowersupply,
swapcablefromunresponsive
actuatorwithafunctional
actuator .Reconnectpower
andtestfunctions.Faulty
actuatorwillnotwork
regardlessoftheconnection
port.
ContactInvacareContinuing
CareServiceDepartment
1-800-668-2337.
Actuatornotworking.Faultyhandcontrolpendant.Pendantcableisconnected,
powerison,lock-out
isoffandhandcontrol
pendantdoesnotfunction.
Disconnectpowersupply
thenreplacehandpendant
withfunctioningunit(from
anotherbed,ifavailable).
Reconnectpowerandtest.
ContactInvacareContinuing
CareServiceDepartment
1-800-668-2337.
Bedendstonottproperly.Bentbrackets.Visuallyensurethebrackets
areparalleltotheframe.
Carefullystraightenminor
deformations.CallService
Departmenttoreplace
signicantlybentbrackets,
1-800-668-2337.