InMode Without Vasculaze Operator Manual
Section 6 Diolaze Treatment
Information
41 of 96
Set treatment parameters. Typical treatment parameters are shown in the
table here:
Skin
Type
]
2
Fluence [J/cm
Pulse
Width
Cooling
Max.
Speed
I
30-40
Test - Wait at least
15-30 min.
increase or decrease
2
5 J/cm-by 2
Short
Strong/Medium/
Weak
Fast
II
25-40
Test - Wait at least
15-30 min.
increase or decrease
2
5 J/cm-by 2
Short
Strong/Medium
/Weak
Fast
III
20-40
Test - Wait at least
15-30 min.
increase or decrease
2
5 J/cm-by 2
Short
Strong/Medium
Fast
IV
20-35
20-40
Test - Wait at least
15-30 min.
increase or decrease
2
3 J/cm-by 2
Short
Long
Strong/Medium
Fast
V
15-35
Test - Wait at least 2-
3 days.
increase or decrease
2
3 J/cm-by 2
Long
Strong
Fast/N
ormal
VI
15-25
Test - Wait at least 2-
3 days.
Increase or decrease
2
2 J/cm-by 1
Long
Strong
Normal
For
longer
cooling
Always start with a low settings level to check patient tolerance, test, watch
epidermal response, and increase or decrease settings gradually. Darker skin is
safer to treat with long pulse width. Lighter hair requires higher fluence to